test that be carried out under the following conditions: tested as table-top equipment, induction coil as of EN 61000-4-8. immersion method, the current flowing through the coil shall be applied in the most unfavourable conditions of phase and direction compared to the voltage(s) energising the meter, frequency equal to reference frequency, magnetic field applied in three perpendicular planes, meter in operating condition, voltage and auxiliary circuits energised with reference voltage, current and power factor as specified in the relevant standard, test with continuous field, field strength 0,5 mT, test/katse, mis tuleb läbi viia alljärgnevatel tingimustel: lauapealne seade, induktsioonpool vastavalt standardile EN 61000-4-8, sukeldusmeetod, pooli läbiv vool peab olema rakendatud arvesti ühenduspinge(te) suhtes kõige ebasobivamatel suundadel ja faasinurkadel, sagedus on võrdne normaalsagedusega, magnetväli rakendatakse kolme ristiasetatud telje suunal, tööolukorras arvesti, pingeahelad ja abiahelad pingestatakse normaalpingega, vool ja võimsustegur on määratud asjakohase standardiga, pidevalt ühtlane magnetväli, väljatihedusega 0,5 mT