The irradiance produced by the power input which the initial calibration has shown to correspond to an irradiance of 3 W/cm? shall be frequently checked (at least once every 50 operating hours) and the apparatus shall be recalibrated if such a check reveals a deviation greater than 0,06 W/cm?.