test that be carried out under the following conditions: table-top equipment, meter in operating condition voltage and auxiliary circuits energized with reference voltage, current and power factor as specified in the relevant standard, cable length between coupling and EUT 1 m, the test voltage shall be applied in common mode in turn to the voltage circuits,the current circuits, if separated from the voltage circuits in normal operation, the auxiliary circuits, if separated from the voltage circuits in normal operation, test voltage on the current and voltage circuits 4 kV, test voltage on the auxiliary circuits with a reference voltage over 40 V to 2 kV, repetition rate 5 kHz, and duration of the test 60 s at each polarity, test/katse, mis viiakse läbi alljärgevatel tingimustel: lauapealne seade, tööolukorras arvesti, pingeahelad ja abiahelad pingestatakse normaalpingega, vool ja võimsustegur on määratud asjakohase standardiga, sidestusseadme ja katseobjekti vahelise kaabli pikkus 1 m, pinge tuleb rakendada assümmeetrilise sisestusviisiga pingeahelatele ja vooluahelatele, kui need tavakasutuses on pingeahelast eraldatud, abiahelatele, kui need on tavakasutuses pingeahelatest eraldatud, katsepinge vooluahelatele ja pingeahelatele 4 kV, katsepinge abiahelatele normaalpingega üle 40 V kuni 2 kV, kordussagedus 5 kHz, katse pikkus 60 s mõlema polaarsusega